Accuracy of drug store blood pressure monitors: an observational study
Katherine L Ross1, Sanchit Bhasin, Merne P Wilson
1Department of Medicine, University of Saskatchewan, Saskatoon, Saskatchewan, Canada.
Objectives:
Publically accessible blood pressure monitors are widely used, but little information is available on their accuracy. We compared blood pressure readings of 17 drug store monitors with those obtained using a validated home monitor (Omron BP742CAN) and both with those taken at home using the Canadian Hypertension Education Program protocol.
Materials And Methods:
Duplicate readings were taken using the drug store monitor (VitaStat, n=6, and PharmaSmart, n=11) on the left arm and the Omron on the right in three participants: two normal and one untreated hypertensive patient. We used Bland-Altman methods for comparison. We explored the correlation with average home blood pressure readings.
Results:
Home average blood pressure for our three participants was 121±6/73±5, 106±6/62±4, and 142±8/81±7 mmHg. The mean systolic blood pressure difference (drug store-Omron) was -1.8±8.2 mmHg. Diastolic pressure difference was 1.7±5.6. Individual paired systolic differences varied from -19 to 14 mmHg. For the participant who required a large cuff, drug store systolic readings tended to be higher (4.1±6.7). In our three participants, drug store monitors as a group read higher than home systolic blood pressure: 7.5 [95% confidence interval (CI) 1.5-13.4], 1.2 (95% CI -4.0 to 6.4), and 1.0 (95% CI -2.5 to 4.4) mmHg. Diastolic blood pressure and heart rate differences were similar in magnitude.
Conclusion:
On average, drug store monitors recorded lower systolic blood pressures and higher diastolic blood pressures than a validated monitor, but the difference was neither statistically nor clinically significant. Single reading comparisons showed a much broader range. In three participants, drug store monitors did reflect the average home blood pressure.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement