Bipolar-power-transistor-based limiter for high frequency ultrasound imaging systems
Hojong Choi1, Hao-Chung Yang1, K Kirk Shung1
1NIH Transducer Resource Center and Department of Biomedical Engineering, University of Southern California, Los Angeles, CA, USA.
Ultrasonics
|November 9, 2013
Summary
A new bipolar power transistor (BPT) limiter offers superior performance for high frequency ultrasound imaging. This BPT limiter significantly reduces signal attenuation, response time, and harmonic distortion compared to traditional methods.
Area of Science:
- Medical Imaging
- Electrical Engineering
- Acoustics
Background:
- High frequency ultrasound imaging systems require protection for receivers from high voltage (HV) spikes.
- Traditional diode limiters and commercially available options have limitations in performance at high frequencies.
Purpose of the Study:
- To introduce and evaluate a novel bipolar power transistor (BPT) limiter configuration.
- To compare the performance of the BPT limiter against traditional diode limiters and commercial limiters.
- To assess limiter performance based on insertion loss (IL), total harmonic distortion (THD), and response time (RT).
Main Methods:
- A new bipolar power transistor (BPT) limiter was designed and fabricated.
- The BPT limiter's performance was evaluated for IL, THD, and RT at 100 MHz.
- A comparative analysis was conducted against a diode limiter and a commercial limiter.
- A two-way pulse-echo test was performed using a 100 MHz single element transducer connected to each limiter.
Main Results:
- The BPT-based limiter demonstrated lower insertion loss (-7.7 dB), total harmonic distortion (-74.6 dB), and a faster response time (43 ns) at 100 MHz compared to the commercial limiter.
- Transducer evaluation showed the BPT-based limiter improved -6 dB bandwidth by 22% and sensitivity by 140% over the commercial limiter.
- The BPT limiter effectively minimized signal attenuation, response time, and total harmonic distortion at high frequencies.
Conclusions:
- The BPT-based limiter is a high-performance solution for high frequency ultrasound imaging systems.
- It offers significant advantages over commercial limiters in terms of signal integrity and speed.
- The BPT limiter's design is well-suited for demanding high frequency ultrasound applications.