Hounsfield units variations: impact on CT-density based conversion tables and their effects on dose distribution
B Zurl1, R Tiefling, P Winkler
1Department of Therapeutic Radiotherapy and Oncology, University Medical School of Graz, Comprehensive Cancer Center Graz, Auenbruggerplatz 32, 8036, Graz, Austria, brigitte.zurl@klinikum-graz.at.
Purpose:
Determination of dose error margins in radiation therapy planning due to variations in Hounsfield Units (HU) values dependent on the use of different CT scanning protocols.
Patients And Methods:
Based on a series of different CT scanning protocols used in clinical practice, conversion tables for radiation dose calculations were generated and subsequently tested on a phantom. These tables were then used to recalculate the radiation therapy plans of 28 real patients after an incorrect scanning protocol had inadvertently been used for these patients.
Results:
Different CT parameter settings resulted in errors of HU values of up to 2.6% for densities of <1.1 g/cm(3), but up to 25.6% for densities of >1.1 g/cm(3). The largest errors were associated with changes in the tube voltage. Tests on a virtual water phantom with layers of variable thickness and density revealed a sawtooth-shaped curve for the increase of dose differences from 0.3 to 0.6% and 1.5% at layer thicknesses of 1, 3, and 7 cm, respectively. Use of a beam hardening filter resulted in a reference dose difference of 0.6% in response to a density change of 5%. The recalculation of data from 28 patients who received radiation therapy to the head revealed an overdose of 1.3 ± 0.4% to the bone and 0.7 ± 0.1% to brain tissue. On average, therefore, one monitor unit (range 0-3 MU) per 100 MU more than the correct dose had been given.
Conclusion:
Use of different CT scanning protocols leads to variations of up to 20% in the HU values. This can result in a mean systematic dose error of 1.5%. Specific conversion tables and automatic CT scanning protocol recognition could reduce dose errors of these types.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Biological Effects of Radiation
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Dose Response Curve: Conventional Versus Nonmonotonic
Dose Size and Dosing Frequency: Determination Methods