Related Experiment Video
Updated: May 6, 2026

Implantation of Left Ventricular Assist Device (LVAD) in Juvenile Landrace Swine: A LVAD Implantation Model of Pediatric Heart Failure
Published on: January 16, 2026
Collateral approach for LV lead implantation in a case with abnormal venous anatomy
Harinder K Bali1, Kapil K Chattree, Surinder K Bali
1Director Cardiology, Dept. of Cardiology, Fortis Hospital, Sector 62, Phase VIII, Mohali 160062, Punjab, India.
Abstract:
A 75-year-old man, 8 years after CABG, with ischemic cardiomyopathy underwent cardiac resynchronization therapy (CRT) for refractory heart failure. Retrograde occlusion venography revealed absence of lateral vein. A functionally occluded middle cardiac vein with branch to anterolateral vein was used for left ventricular lead implantation. Using a collateral route for left ventricular lead implantation is a new technique. Lead position was stable with excellent threshold. Follow-up at 6 months reveals continued stable lead position.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
06:04Upper-extremity Approach for Secondary Access in Transfemoral Transcatheter Aortic Valve Implantation
Published on: August 8, 2025