Related Experiment Video
Updated: May 6, 2026

Author Spotlight: Advancements in Impedance Monitoring for Cochlear Implant Surgery
Published on: August 4, 2023
In vivo evaluation of mastication noise reduction for dual channel implantable microphone
SeongTak Woo1, EuiSung Jung, HyungGyu Lim
1Graduate School of Electrical Engineeing, Kyungpook National University, Sangyuk-dong, Buk-Gu, Daegu, 702-701, S. Korea.
Abstract:
Input for fully implantable hearing devices (FIHDs) is provided by an implantable microphone under the skin of the temporal bone. However, the implanted microphone can be affected when the FIHDs user chews. In this paper, a dual implantable microphone was designed that can filter out the noise from mastication. For the in vivo experiment, a fabricated microphone was implanted in a rabbit. Pure-tone sounds of 1 kHz through a standard speaker were applied to the rabbit, which was given food simultaneously. To evaluate noise reduction, the measured signals were processed using a MATLAB program based adaptive filter. To verify the proposed method, the correlation coefficients and signal to-noise ratio before and after signal processing were calculated. By comparing the results, signal-to-noise ratio and correlation coefficients are enhanced by 6.07dB and 0.529 respectively.
More Related Videos
06:04Systematic Hearing Performance Evaluation Process for Adolescents with Cochlear Implantation at Early Ages
Published on: March 24, 2023
08:42The DREAM Implant: A Lightweight, Modular, and Cost-Effective Implant System for Chronic Electrophysiology in Head-Fixed and Freely Behaving Mice
Published on: July 26, 2024