Purpose:
Perioperative hypothermia is still a common occurrence, and it can be difficult to measure a patient's core temperature accurately, especially during regional anesthesia, with placement of a laryngeal mask airway device, or postoperatively. We evaluated a new disposable double-sensor thermometer and compared the resulting temperatures with those of a distal esophageal thermometer and a bladder thermometer in patients undergoing general and regional anesthesia, respectively. Furthermore, we compared the accuracy of the thermometer between regional and general anesthesia, since forehead microcirculation might differ between the two types of anesthesia.
Methods:
We assessed core temperature in 36 general anesthesia patients and 20 patients having regional anesthesia for orthopedic surgery. The temperatures obtained using the double-sensor thermometer were compared with those obtained with the distal esophageal thermometer in the general anesthesia population and those obtained with the bladder thermometer in regional anesthesia patients.
Results:
In our general anesthesia patients, 90% (95% confidence interval [CI] 85 to 95) of all double-sensor values were within 0.5°C of esophageal temperatures. The average difference (bias) between the esophageal and double-sensor temperatures was -0.01°C. In patients undergoing regional anesthesia 89% (95% CI 80 to 97) of all double-sensor values were within 0.5°C of bladder temperatures. The average difference (bias) between the bladder and double-sensor temperatures was -0.13°C, limits of agreement were -0.65 to 0.40°C.
Conclusions:
In a perioperative patient population undergoing general or regional anesthesia, the accuracy of the noninvasive disposable double-sensor thermometer is comparable with that of the distal esophageal and bladder thermometers in routine clinical practice. Furthermore, the sensor performed comparably in patients undergoing regional and general anesthesia.
Related Concept Videos
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...