Related Concept Videos

Amperometry: Overview01:10

Amperometry: Overview

2.2K
Amperometry is a technique commonly used to measure the concentration of specific analytes in a solution by monitoring the electric current generated during an electrochemical reaction. It involves applying a constant potential between a working electrode and a reference electrode to measure the resulting current, which is proportional to the concentration of the analyte. The Clark oxygen electrode operates based on this principle of amperometry. It consists of a cathode and an anode enclosed...
2.2K
Potentiometry: Membrane Electrodes01:15

Potentiometry: Membrane Electrodes

2.3K
Membrane electrodes, also known as p-ion electrodes, use membranes that selectively interact with free analyte ions, generating a potential difference across the membrane. The resulting membrane potential, known as the asymmetry potential, is not zero even when analyte concentrations on both sides of the membrane are equal. The membrane's response is typically not selective to a single analyte but proportional to the concentration of all ions in the sample solution capable of interacting at...
2.3K
Microbial Biosensors01:17

Microbial Biosensors

91
Microbial biosensors are analytical devices that utilize living microbes to detect specific substances through measurable signals. These devices consist of two main components: biosensing organisms and signal-transducing elements. Biosensing organisms, such as Escherichia coli or Saccharomyces cerevisiae, are typically housed in multiwell plates connected to transducers, enabling rapid, real-time detection of target analytes.Signal Generation MechanismWhen a target analyte—such as...
91

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Comparative clinical outcomes of suprapatellar intramedullary nailing vs. minimally invasive plate osteosynthesis fixation for distal tibial fractures: a retrospective cohort study.

Frontiers in surgery·2026
Same author

Impact on clinical biomarkers, immune recovery, and HIV reservoir dynamics of switching to EVG/c/FTC/TAF versus maintaining a non-INSTI-based regimen in virologically suppressed individuals.

Infectious diseases & immunity·2026
Same author

Safety Profile of COVID-19 Vaccines in HIV Patients Undergoing ART and Their Impact on Immune Recovery and HIV Reservoirs.

Infectious diseases & immunity·2026
Same author

Multi-omics unravels testis-regulated genetic mechanism of musk secretion in muskrat (Ondatra zibethicus).

BMC genomics·2026
Same author

Preparation of antimicrobial biopolymer particles through co-precipitation of hen egg white lysozyme and poly(hydroxy butyrate).

International journal of biological macromolecules·2026
Same author

Validating the accuracy of the YuWell YE680B automated upper-arm blood pressure monitor in pediatric and adult populations according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (ISO 81060-2:2018/Amd.1:2020/Amd.2:2024).

Blood pressure monitoring·2026
Same journal

Microfluidic rare cell analysis beyond counting: workflow design from enrichment to multi-omics.

Lab on a chip·2026
Same journal

A sperm racetrack to separate sperm by swim speed.

Lab on a chip·2026
Same journal

Controlled encapsulation and droplet size prediction in two-step microfluidic double emulsions.

Lab on a chip·2026
Same journal

A particulate blood-mimicking fluid with physiological biconcave geometry for microscale hemorheology.

Lab on a chip·2026
Same journal

Multicellular sensor arrays fabricated by capillary stamping for pattern-based odor discrimination.

Lab on a chip·2026
Same journal

A real-time microfluidic surveillance system for multiplex detection of heavy metal contamination in wastewater.

Lab on a chip·2026
See all related articles

Related Experiment Video

Updated: May 6, 2026

Hollow Microneedle-based Sensor for Multiplexed Transdermal Electrochemical Sensing
08:19

Hollow Microneedle-based Sensor for Multiplexed Transdermal Electrochemical Sensing

Published on: June 1, 2012

13.7K

A differential dielectric affinity glucose sensor.

Xian Huang1, Charles Leduc, Yann Ravussin

  • 1Department of Mechanical Engineering, Columbia University, New York, NY 10027, USA. qlin@columbia.edu.

Lab on a Chip
|November 14, 2013
PubMed
Summary
This summary is machine-generated.

A novel continuous glucose monitor uses a differential dielectric sensor for accurate glucose tracking in interstitial fluid. This implantable device offers improved stability and precision for potential clinical use in diabetic patients.

More Related Videos

NiO Nanoflowers for Non-Enzymatic Amperometric Detection of Glucose
11:04

NiO Nanoflowers for Non-Enzymatic Amperometric Detection of Glucose

Published on: December 30, 2025

307
Measuring Glucose Uptake in Drosophila Models of TDP-43 Proteinopathy
07:07

Measuring Glucose Uptake in Drosophila Models of TDP-43 Proteinopathy

Published on: August 3, 2021

2.4K

Related Experiment Videos

Last Updated: May 6, 2026

Hollow Microneedle-based Sensor for Multiplexed Transdermal Electrochemical Sensing
08:19

Hollow Microneedle-based Sensor for Multiplexed Transdermal Electrochemical Sensing

Published on: June 1, 2012

13.7K
NiO Nanoflowers for Non-Enzymatic Amperometric Detection of Glucose
11:04

NiO Nanoflowers for Non-Enzymatic Amperometric Detection of Glucose

Published on: December 30, 2025

307
Measuring Glucose Uptake in Drosophila Models of TDP-43 Proteinopathy
07:07

Measuring Glucose Uptake in Drosophila Models of TDP-43 Proteinopathy

Published on: August 3, 2021

2.4K

Area of Science:

  • Biomedical Engineering
  • Sensor Technology
  • Diabetes Management

Background:

  • Continuous glucose monitoring (CGM) is crucial for diabetes management.
  • Existing CGM technologies face challenges with accuracy, stability, and invasiveness.
  • A need exists for reliable, implantable glucose sensors for real-time monitoring.

Purpose of the Study:

  • To present a novel continuous glucose monitor (CGM) utilizing a differential dielectric sensor.
  • To evaluate the performance and stability of the sensor for interstitial fluid glucose measurement.
  • To assess the potential clinical applicability of the device for diabetic patients.

Main Methods:

  • Development of a differential dielectric sensor using microelectromechanical systems (MEMS) technology.
  • Implantation of sensing and reference modules with glucose-specific and reference solutions within microchambers.
  • Differential measurement of permittivity changes to determine glucose concentration in interstitial fluid.

Main Results:

  • The sensor accurately measured glucose concentrations from 0 to 500 mg/dL in vitro.
  • Achieved high resolution (~1.7 μg/dL) and accuracy (~1.74 mg/dL).
  • Demonstrated significant improvements in device stability, reducing drift in uncontrolled and temperature-varying environments.

Conclusions:

  • The differential dielectric sensor offers a stable and accurate method for continuous glucose monitoring.
  • Preliminary animal studies confirm its ability to track blood glucose levels.
  • The developed CGM holds promise for clinical application in managing diabetes.