Related Experiment Video
Updated: May 6, 2026

Flexible Colonoscopy in Mice to Evaluate the Severity of Colitis and Colorectal Tumors Using a Validated Endoscopic Scoring System
Published on: October 16, 2013
Magnetic endoscopic imaging vs standard colonoscopy: meta-analysis of randomized controlled trials
Yi Chen1, Yu-Ting Duan, Qin Xie
1Yi Chen, Bo Chen, Lin Xia, Xian-Peng Qin, Yong Zhou, Ning-Ning Li, Xiao-Ting Wu, Department of Gastrointestinal Surgery, West China Hospital, Sichuan University, Chengdu 610041, Sichuan Province, China.
Aim:
To assess the theoretical advantages of magnetic endoscope imaging (MEI) over standard colonoscopies (SCs) and to compare their efficacies.
Methods:
Electronic databases, including PubMed, EMBASE, the Cochrane library and the Science Citation Index, were searched to retrieve relevant trials. In addition, abstracts from papers presented at professional meetings and the reference lists of retrieved articles were reviewed to identify additional studies. The meta-analyses were performed using RevMan 5.1. A random effect model with the Mantel-Haenszel method was used for pooling dichotomous and continuous data. A sensitivity analysis was performed by excluding the trials with a small number of patients and by excluding the trials performed by inexperienced providers.
Results:
Eight randomized controlled trials (RCTs), including 2967 patients, were included in the meta-analysis to compare cecal intubation rates and times, sedation dose, abdominal pain scores and the use of ancillary maneuvers between MEI and SC. The overall OR was 1.92 (95%CI: 1.13-3.27, eight RCTs), as indicated by the cecal intubation rate of MEI compared with SC, but MEI did not have any distinct advantage over SC for cecal intubation time (MD = -0.07, 95%CI: -0.16-0.02; three RCTs). MEI did not generally result in lower pain scores. Outcomes were also analyzed for the two subgroups based on the endoscopists' experience level to evaluate cecal intubation rates. MEI presented better outcomes for non-experienced colonoscopists than experienced colonoscopists.
Conclusion:
The real-time magnetic imaging system is of benefit in training and educating inexperienced endoscopists and improves the cecal intubation rate for experienced and inexperienced endoscopists.
More Related Videos
09:01Non-invasive Assessment of the Efficacy of New Therapeutics for Intestinal Pathologies Using Serial Endoscopic Imaging of Live Mice
Published on: March 10, 2015
09:42Murine Endoscopy for In Vivo Multimodal Imaging of Carcinogenesis and Assessment of Intestinal Wound Healing and Inflammation
Published on: August 26, 2014
Related Concept Videos
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Endoscopic Procedures II: Colonoscopy
Endoscopic Procedures III: Video Capsule Endoscopy
Endoscopic Procedures V: ERCP
Patient...
Magnetic Resonance Imaging
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...