Comparison of invasive and oscillometric blood pressure measurement techniques in anesthetized sheep, goats, and
Turi K Aarnes1, John Ae Hubbell, Phillip Lerche
1Department of Veterinary Clinical Sciences, College of Veterinary Medicine, The Ohio State University, Columbus, OH, USA.
Objective:
To determine the level of agreement between an oscillometric (O-NIBP) and an invasive method (IBP) of monitoring arterial blood pressure (ABP) in anesthetized sheep, goats, and cattle.
Study Design:
Prospective clinical study.
Animals:
Twenty sheep and goats, 20 cattle weighing < 150 kg body weight, and 20 cattle weighing 150 kg body weight.
Methods:
Animals were anesthetized and systolic ABP (SABP), mean ABP (MABP), and diastolic ABP (DABP) were measured using IBP and O-NIBP. Differences between IBP and O-NIBP, and 95% limits of agreement (LOA) between SABP, MABP, and DABP values were assessed by the Bland-Altman method.
Results:
Mean difference ± standard deviation (range) between SABP, DABP, and MABP measurements in sheep and goats was 0 ± 16 (-57 to 38) mmHg, 13 ± 16 (-37 to 70) mmHg, and 8 ± 13 (-34 to 54) mmHg, respectively. Mean difference between SABP, DABP, and MABP measurements in small cattle was 0 ± 19 (-37 to 37) mmHg, 6 ± 18 (-77 to 48) mmHg, and 4 ± 16 (-73 to 48) mmHg, respectively. Mean difference between SABP, DABP, and MABP measurements in large cattle was -18 ± 32 (-107 to 71) mmHg, 7 ± 29 (-112 to 63) mmHg, and -5 ± 28 (-110 to 60) mmHg, respectively. The 95% LOAs for SABP, DABP, and MABP were -31 to +31, -19 to +44, and -19 to +34 mmHg, respectively in sheep and goats; were -37 to +37, -19 to +44, and -19 to +34 mmHg, respectively in small cattle; and were -81 to +45, -50 to +63, and -59 to +50 mmHg, respectively in large cattle.
Conclusions:
Agreement was poor between O-NIBP and IBP monitoring techniques.
Clinical Relevance:
Arterial BP should be monitored in anesthetized sheep, goats, and cattle using IBP.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure