Unidirectional valve malfunction by the breakage or malposition of disc - two cases report -
Chol Lee1, Kyu Chang Lee, Hye Young Kim
1Department of Anesthesiology and Pain Medicine, Wonkwang University College of Medicine, Iksan, Korea.
Abstract:
Malfunction of the unidirectional valve in a breathing circuit system may cause hypercapnia from the rebreathing of expired gas, ventilation failure, and barotrauma. Capnography is a useful method for monitoring the integrity of the unidirectional valve. We experienced two cases of malfunction of a unidirectional valve which caused leakage and reverse flow, diagnosed early as a change of the capnographic waveform. One case was caused by expiratory unidirectional valve breakage. The other was caused by an incorrectly-assembled inspiratory unidirectional valve.
More Related Videos
06:45Design and Application of a Fault Detection Method Based on Adaptive Filters and Rotational Speed Estimation for an Electro-Hydrostatic Actuator
Published on: October 28, 2022
07:03Author Spotlight: Integrating Computational and Experimental Approaches in Precision Oncology
Published on: December 1, 2023
Related Concept Videos
Degenerative Disc Disease ll: Pathophysiology
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Herniated Intervertebral Disc l: Introduction
Directional Relays
Aortic Regurgitation II: Clinical Features and Diagnostic Tests
Degenerative Disc Disease I: Introduction