Inductively coupled plasma mass spectrometry with an enlarged sampling orifice and offset Ion Lens. I. Ion
1Ames Laboratory-U.S. Department of Energy, Department of Chemistry, Iowa State University, 50011, Ames, IA, USA.
Abstract:
A new inductively coupled plasma mass spectrometer (ICP-MS) with four stages of differential pumping is described. The relatively large sampling orifice (1.31-mm dia.) improves signals for metal ions and resists plugging from deposited solids. A new ion lens is described that deflects ions off center and then back on center into the differential pumping orifice; there is no photon stop in the center of the beam. Calculations of ion trajectories using SIMIOn show that only those ions that leave the skimmer on center are transmitted, whereas most other lenses used in ICP-MS transmit only ions that leave the skimmer off axis. The performance of a Channeltron electron multiplier is compared to that of a Daly detector. Both detectors yield similar sensitivities of ≈ 10(6) counts s(-1) per ppm and detection limits of ≈ 1 pptr. The background with a Channeltron electron multiplier is only 0.4 counts s(-1) and is only slightly higher than the dark current count rate. Presumably the offset ion lens used in the present work efficiently screens the detector from photons emitted by the plasma. The background with the Daly detector is 4 counts s(-1), which represents a substantial improvement over the background obtained in previous use of the Daly detector with JCP-MS.
Related Concept Videos
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Mass Analyzers: Common Types
Mass Spectrometers