Preparing an Isotopically Pure 229Th Ion Beam for Studies of 229mTh
Published on: May 3, 2019
Coupling a particle beam interface directly to a quadrupole ion trap mass spectrometer
M E Bier1, P C Winkler, J R Herron
1Finnigan MAT Corporation, 355 River Oaks Parkway, 95134, San Jose, CA, USA.
A new particle beam interface coupled with a quadrupole ion trap mass spectrometer enables direct analysis of analytes in solution. This system efficiently collects electron ionization mass spectra, detecting caffeine at low levels.
Area of Science:
- Analytical Chemistry
- Mass Spectrometry
- Instrumentation
Background:
- Traditional mass spectrometry often requires extensive sample preparation.
- Direct analysis of analytes in solution presents challenges in sample introduction and ionization.
- Coupling particle beam interfaces with mass spectrometers can improve sample throughput.
Purpose of the Study:
- To develop and evaluate a novel particle beam interface for quadrupole ion trap mass spectrometry.
- To enable direct collection of electron ionization mass spectra from analytes in solution.
- To assess the performance and detection limits of the coupled system.
Main Methods:
- A particle beam interface was coupled to a quadrupole ion trap mass spectrometer.
- The interface included a pneumatic nebulizer, heated desolvation chamber, and three-stage separator.
- Mass analysis utilized mass-selective instability scan with resonance ejection.
Main Results:
- Electron ionization mass spectra were successfully obtained for caffeine, carbaryl, 17α-hydroxyprogesterone, and reserpine.
- Caffeine was detected at 25 pmol (5 ng) with a signal-to-noise ratio of 50.
- Nonlinear caffeine calibration was observed below 30 ng, with a more linear response at higher concentrations.
Conclusions:
- The coupled system provides a viable method for direct analysis of analytes in solution via electron ionization mass spectrometry.
- The interface effectively separates solvent molecules and transfers particles for ionization.
- The system demonstrates good sensitivity for detecting trace amounts of analytes.
More Related Videos
Related Concept Videos
Mass Analyzers: Common Types
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Tandem Mass Spectrometry
Mass Spectrometers
Chemical Ionization (CI) Mass Spectrometry
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...