A Practical Guide on Coupling a Scanning Mobility Sizer and Inductively Coupled Plasma Mass Spectrometer SMPS-ICPMS
Published on: July 11, 2017
Selection of solvent load and first-stage pressure to reduce interference effects in inductively coupled plasma-mass
1Department of Chemistry, Indiana University, 47405, Bloomington, IN.
Abstract:
In inductively coupled plasma-mass spectrometry the first-stage pressure and solvent characteristics can strongly influence spectral and nonspectroscopic interference effects. By manipulating the pressure and solvent load, one can regulate the degree of analyte signal suppression observed in the presence of high concentrations (> 10 mM) of concomitants. Importantly, the same operating conditions that eliminate the matrix effects maintain the analytical utility of the system. However, for some interferent-analyte combinations, the identity of the concomitant anion and subsequent pH of the solution determine whether the interference effects can be eliminated entirely. The first-stage pressure does not appear to significantly affect the oxide-ion and doubly charged ion ratios; the solvent characteristics are the dominant factors that dictate these ratios.
Related Concept Videos
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Atomic Emission Spectroscopy: Interference
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
High-Performance Liquid Chromatography: Elution Process