A comparison of two nasal continuous positive airway pressure interfaces--a randomized crossover study
Background:
In spite of widespread use of nasal CPAP there are comparatively few studies to guide the choice of nasal prongs.
Objectives:
To determine whether the Fisher & Paykel Healthcare (FPH) neonatal continuous positive airway pressure (CPAP) interface was effective in providing bubble CPAP when compared to the Hudson prong interface.
Methods:
The study was a randomized cross-over study of twenty newborn infants 500 g or more requiring CPAP for respiratory support at birth. Infants were randomized to either the Fisher & Paykel Healthcare or Hudson CPAP interface for twenty four hours. Crossover between interfaces occurred after subsequent twenty four hour periods. The primary outcome was the provision of desired CPAP pressures, defined as provision of CPAP within ± one cm H2O of set pressure.
Results:
The percentage time CPAP was within ± one cm H2O of set pressure was 66.5% for the Hudson and 71.8% for the FPH interface (p = 0.66). Oxygen saturations for the Hudson interface were in target range for a median of 97.8% of the time, and, with the FPH interface, for a median of 98.2% of the time (p = 0.76). Clinically significant differences in primary or secondary outcomes between the two groups were not detected.
Conclusions:
The nasal CPAP interfaces studied were equally effective in achieving desired bubble CPAP pressures and target saturations.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Crossover Experiments
Crossover designs are performed even with smaller sample sizes since the samples can act as their controls. These are better than simple randomized trials since patients are exposed to all the treatments.
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Suctioning the Nasopharyngeal Airway
Equipment Required
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...