A comparison of two nasal continuous positive airway pressure interfaces--a randomized crossover study
This study found that both Fisher & Paykel Healthcare and Hudson nasal prong interfaces effectively deliver bubble continuous positive airway pressure (CPAP) in newborns. No significant differences were observed in achieving target pressures or oxygen saturations between the two CPAP interfaces.
Area of Science:
- Neonatal respiratory support
- Medical device comparison
- Pulmonary medicine
Background:
- Limited research guides the selection of nasal prongs for neonatal continuous positive airway pressure (CPAP).
- Nasal CPAP is widely used for respiratory support in newborns.
Purpose of the Study:
- To compare the effectiveness of the Fisher & Paykel Healthcare (FPH) neonatal CPAP interface against the Hudson prong interface.
- To evaluate the ability of each interface to deliver bubble CPAP accurately.
Main Methods:
- A randomized cross-over study involving twenty newborn infants requiring CPAP.
- Infants received either the FPH or Hudson interface for 24-hour periods, with crossover between devices.
- The primary outcome measured was the delivery of CPAP within ±1 cm H2O of the set pressure.
Main Results:
- The FPH interface delivered CPAP within the target pressure range 71.8% of the time, compared to 66.5% for the Hudson interface (p=0.66).
- Median oxygen saturation in the target range was 98.2% for FPH and 97.8% for Hudson (p=0.76).
- No clinically significant differences were detected in primary or secondary outcomes between the two interfaces.
Conclusions:
- Both the Fisher & Paykel Healthcare and Hudson nasal CPAP interfaces demonstrated comparable effectiveness.
- The studied interfaces equally achieved desired bubble CPAP pressures and maintained target oxygen saturations in neonates.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Crossover Experiments
Crossover designs are performed even with smaller sample sizes since the samples can act as their controls. These are better than simple randomized trials since patients are exposed to all the treatments.
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Suctioning the Nasopharyngeal Airway
Equipment Required
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...