Reliability of simulated robustness testing in fast liquid chromatography, using state-of-the-art column technology,
Róbert Kormány1, Jenő Fekete2, Davy Guillarme3
1Egis Pharmaceuticals Plc. , Keresztúri út 30-38, 1106, Budapest, Hungary.
Abstract:
The goal of this study was to evaluate the accuracy of simulated robustness testing using commercial modelling software (DryLab) and state-of-the-art stationary phases. For this purpose, a mixture of amlodipine and its seven related impurities was analyzed on short narrow bore columns (50×2.1mm, packed with sub-2μm particles) providing short analysis times. The performance of commercial modelling software for robustness testing was systematically compared to experimental measurements and DoE based predictions. We have demonstrated that the reliability of predictions was good, since the predicted retention times and resolutions were in good agreement with the experimental ones at the edges of the design space. In average, the retention time relative errors were <1.0%, while the predicted critical resolution errors were comprised between 6.9 and 17.2%. Because the simulated robustness testing requires significantly less experimental work than the DoE based predictions, we think that robustness could now be investigated in the early stage of method development. Moreover, the column interchangeability, which is also an important part of robustness testing, was investigated considering five different C8 and C18 columns packed with sub-2μm particles. Again, thanks to modelling software, we proved that the separation was feasible on all columns within the same analysis time (less than 4min), by proper adjustments of variables.
More Related Videos
10:21Automated Hydrophobic Interaction Chromatography Column Selection for Use in Protein Purification
Published on: September 21, 2011
14:42Liquid Chromatography Coupled to Refractive Index or Mass Spectrometric Detection for Metabolite Profiling in Lysate-based Cell-free Systems
Published on: September 23, 2021
Related Concept Videos
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Principles Of Column Chromatography
Supercritical Fluid Chromatography
SFC utilizes a supercritical fluid mobile phase,...
High-Performance Liquid Chromatography: Elution Process
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...