Variability of M-mode versus two-dimensional echocardiography measurements in children with dilated cardiomyopathy
Caroline K Lee1, Renee Margossian, Lynn A Sleeper
1Washington University in St. Louis, St. Louis, MO, USA, lee_c@kids.wustl.edu.
Insights
For pediatric dilated cardiomyopathy (DCM) patients, M-mode echocardiography offers better reproducibility for left-ventricular (LV) dimensions and shortening fraction. Consistent use of either M-mode or 2D imaging is recommended for serial measurements.
Area of Science:
- Pediatric Cardiology
- Echocardiography
- Cardiovascular Imaging
Background:
- M-mode and 2D echocardiography are standard for assessing left-ventricular (LV) size and function in pediatric dilated cardiomyopathy (DCM).
- The comparative reproducibility and agreement between these echocardiographic modalities in pediatric DCM are not well-established.
Purpose of the Study:
- To compare the interreader, intrareader, and interacquisition reproducibility of M-mode versus 2D echocardiographic measurements in pediatric DCM patients.
- To assess the agreement between M-mode and 2D measurements for LV size and function.
Main Methods:
- Prospective, multicenter observational study (Ventricular Volume Variability study).
- Two sonographers performed baseline acquisitions; two readers performed measurements at a core laboratory.
- One reader repeated measurements after 1 month to assess intrareader, interreader, and interacquisition variability.
Main Results:
- M-mode demonstrated similar or superior intrareader and interreader reproducibility for LV dimensions, shortening fraction (SF), and wall thickness.
- 2D imaging showed similar or better interacquisition reproducibility for most variables, except SF.
- Interacquisition variability was approximately double intrareader variability.
- LV dimensions showed high reproducibility and agreement between M-mode and 2D.
Conclusions:
- For serial echocardiographic assessments in pediatric DCM, consistent use of a single method (M-mode or 2D) and reader is recommended to minimize variability.
- M-mode is preferred when SF is the primary measure of interest.
- Except for LV dimensions, M-mode and 2D measurements should not be used interchangeably due to poor agreement.
Abstract:
M-mode and 2-dimensional (2D) echocardiographic imaging are routinely used to quantify left-ventricular (LV) size and function in pediatric patients with dilated cardiomyopathy (DCM). The reproducibility of and correlation between these techniques are unknown. This analysis sought to compare interreader, intrareader, and interacquisition reproducibility of M-mode versus 2D measurements in pediatric DCM patients. The Ventricular Volume Variability study of the Pediatric Heart Network is a multicenter, prospective, observational study assessing the course of chronic DCM in children. Two sonographers performed baseline image acquisitions locally, and two readers performed measurements at the echocardiographic core laboratory. One reader repeated measurements 1 month later. These data were used to assess reproducibility and agreement between M-mode and 2D measurements. One hundred sixty-nine subjects were enrolled. M-mode had similar or greater reproducibility in both intrareader and interreader settings for LV dimensions, shortening fraction (SF), and most wall thicknesses. In contrast, 2D reproducibility was similar or better for nearly all variables in the interacquisition setting but not for SF. Interacquisition variability was approximately twice the intrareader variability. LV dimensions by either modality consistently had high reproducibility and had the highest agreement between modalities. In pediatric DCM patients, variability of linear echocardiographic assessment could be minimized by relying on a single reader and using a consistent method (M-mode or 2D) for serial measurements, preferably M-mode when SF is the primary variable of interest. Except for LV dimensions, M-mode and 2D values should not be used interchangeably due to poor agreement.
More Related Videos
11:50High-frequency High-resolution Echocardiography: First Evidence on Non-invasive Repeated Measure of Myocardial Strain, Contractility, and Mitral Regurgitation in the Ischemia-reperfused Murine Heart
Published on: July 9, 2010
06:34Evaluation of Left Ventricular Structure and Function using 3D Echocardiography
Published on: October 28, 2020
Related Concept Videos
Mitral Stenosis II: Clinical features and Diagnostic Tests
Cardiomyopathy III: Hypertrophic Cardiomyopathy
Cardiomyopathy II: Dilated Cardiomyopathy
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...