70 kV computed tomography of the thorax: valence for computer-assisted nodule evaluation and radiation dose - first
Boris Bodelle1, Dorothée Klement2, Josef Matthias Kerl2
1Clinic of the Goethe University, Department of Diagnostic and Interventional Radiology, Frankfurt, Germany bbodelle@googlemail.com.
Background:
Computed tomography (CT) is the gold standard for evaluation of pulmonary nodules and is at the same time responsible for the majority of the collective effective dose.
Purpose:
To evaluate radiation dose and efficacy of computer-assisted detection (CAD) for solid pulmonary nodules in low dose chest CT performed at 70 kV.
Material And Methods:
CAD was performed upon chest CT with 70 kV and 100 kV (gold standard) at manufacture's recommended tube current of 87 mAs (collimation, 64 × 0.6 mm). Detection rate for pulmonary nodules and size measurements of both techniques were compared to each other. Radiation dosage in terms of effective dose (E) was measured using an Alderson-Rando Phantom.
Results:
Seventy-four patients with 301 solid nodules were included in the study. CAD detection rate was similar for 70 kV (94.7%) and 100 kV (92.4%). Mean transversal nodule diameter was 5.5 mm for 70 kV and 5.7 mm for 100 kV with an average volume of 0.12 mL (both techniques). Derived from the phantom measurements patient examinations resulted in an E of 0.51 mSv (70 kV) versus 2.02 mSv (100 kV).
Conclusion:
70 kV low-dose chest CT is suitable for CAD based lung nodule analysis at a fraction of the radiation burden of the standard technique. Since the measurements are highly accurate, 70 kV CT could be used for detection of pulmonal lesions as well as follow-up studies.
More Related Videos
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT