M H F Schreinemacher1, K W Y van Barneveld, E Peeters
1Department of General Surgery, Maastricht University Medical Centre, PO Box 5800, 6202 AZ, Maastricht, The Netherlands, m.schreinemacher@gmail.com.
Purpose:
Intraperitoneal mesh fixation for hernia repair is associated with adhesion formation. In this experimental study, adhesions against absorbable and non-absorbable fixation methods were compared.
Methods:
Six commercially available fixation methods were placed intraperitoneally in rats with a small pore polypropylene mesh coated on one side with ePTFE (Intramesh T1(®)). Two non-absorbable fixation methods: Prolene(®) (polypropylene) sutures and Protack(®) (titanium) tackers. Four absorbable methods: Vicryl(®) sutures (polyglactin), Absorbatack(®) and Permasorb(®) tackers (both mixes of lactic and glycolic acids) and Tisseel Duo(®) (fibrin glue). Adhesions and histology were studied at 7 and 90 days follow-up. In addition, fixation methods were placed without mesh, in order to study the reaction to the fixation method per se.
Results:
No adhesion formation, but also inadequate mesh fixation was found with Tisseel Duo(®), which had been completely resorbed at 7 days follow-up. Vicryl(®) sutures could no longer be detected at 90 days follow-up and were associated with a favorable adhesion profile. All other fixation methods were still intact 90 days after implantation. When placed without mesh, adhesion formation was significantly less than placed with a mesh (18 vs. 93 %, P < 0.001). Without mesh, adhesions were worst with Permasorb(®) tackers.
Conclusions: