Maintenance of airway pressure during filter exchange due to auto-triggering
Background:
Daily routine ventilator-filter exchange interrupts the integrity of the ventilator circuit. We hypothesized that this might reduce positive airway pressure in mechanically ventilated ICU patients, inducing alveolar collapse and causing impaired oxygenation and compliance of the respiratory system.
Methods:
We studied 40 consecutive ICU subjects (P(aO2)/F(IO2) ratio ≤ 300 mm Hg), mechanically ventilated with pressure-regulated volume control or pressure support and PEEP ≥ 5 cm H2O. Before the filter exchange, (baseline) tidal volume, breathing frequency, end-inspiratory plateau pressure, and PEEP were recorded. Compliance of the respiratory system was calculated; F(IO2), blood pressure, and pulse rate were registered; and P(aO2), P(aCO2), pH, and base excess were measured. Measurements were repeated 15 and 60 min after the filter exchange. In addition, a bench test was performed with a precision test lung with similar compliance and resistance as in the clinical study.
Results:
The exchange of the filter took 3.5 ± 1.2 s (mean ± SD). There was no significant change in P(aO2) (89 ± 16 mm Hg at baseline vs 86 ± 16 mm Hg at 15 min and 88 ± 18 mm Hg at 60 min, P = .24) or in compliance of the respiratory system (41 ± 11 mL/cm H2O at baseline vs 40 ± 12 mL/cm H2O at 15 min and 40 ± 12 mL/cm H2O at 60 min, P = .32). The bench study showed that auto-triggering by the ventilator when disconnecting from the expiratory circuit kept the tracheal pressure above PEEP for at least 3 s with pressure controlled ventilation.
Conclusions:
This study showed that a short disconnection of the expiratory ventilator circuit from the ventilator during filter exchange was not associated with any significant deterioration in lung function 15 and 60 min later. This result may be explained by auto-triggering of the ventilator with high inspiratory flows during the filter exchange, maintaining airway pressure.
More Related Videos
07:15Preoxygenation Techniques for Tracheal Intubation in Critically Ill Adults Utilizing Oxygen Mask and Noninvasive Ventilation
Published on: December 5, 2025
04:16Inspiratory Muscle Training as an Adjunct to the Treatment of Weaning Failure in Critically Ill Patients: A Practical Guide
Published on: January 30, 2026
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....