An automated hand-held elastometer for quantifying the passive stiffness of the levator ani muscle in women
Jennifer A Kruger1, Poul M F Nielsen, Stephanie C Budgett
1Auckland Bioengineering Institute, The University of Auckland, Auckland, New Zealand.
Aim:
Design and develop an automated, hand-held instrument (elastometer) to assess in vivo passive stiffness of the pelvic floor muscle.
Materials And Methods:
The elastometer system consisted of a hand piece, real-time controller, and laptop computer. A cable connected the hand-piece to the controller, which communicated with a laptop computer via an ethernet connection. Force sensitivity calibration and displacement accuracy were determined experimentally using a spring load and an Instron mechanical tester. A test re-test series quantified the in vivo repeatability (within a procedure) and reproducibility (between procedures after a 5 min delay) of passive stiffness in volunteers (n = 20). Stiffness was determined from the gradient of the force-displacement curve for each cycle.
Results:
The force-aperture spring measurements from the elastometer showed consistent (r(2) = 1.0000) agreement with those measured by the Instron. The difference between spring stiffness as measured by the elastometer and the Instron (388.1 N/m cf. 388.5 N/m, respectively) was negligible. The intra-class correlation coefficient for repeatability within procedures was 0.986 95% CI (0.964-0.994) n = 20, and reproducibility between procedures ICC 0.934 (95% CI 0.779-0.981) n = 12. Bland-Altman analysis determined a bias of 0.3 and 18.5 N/m, for repeatability and reproducibility respectively. Neither bias is likely to be clinically significance.
Conclusion:
The elastometer demonstrated very good repeatability and accuracy in the measurement of force/displacement during in vitro testing. There was a high degree of repeatability and reproducibility in stiffness measurements in a test re-test series. Our results demonstrate the elastometer is accurate and reliable and thereby suitable for larger clinical trials.
Related Concept Videos
Assessment of the Rectum and Anus
Rectal Inspection
Begin by inspecting the perianal and anal areas for color, texture, rashes,...
Assessment of the Abdomen III: Palpation
Muscles of the Pelvic Floor and Perineum
Perineal Layer
The perineum is a diamond-shaped area below the pelvic diaphragm, divided into an anterior urogenital triangle that contains the external genitals and a posterior anal triangle housing the anus. The urogenital...
Assessment of the Abdomen II: Percussion
Percussion
Percussion is an essential...