Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025
Coded excitation of ultrasonic guided waves in long bone fracture assessment
Huilin Zhang1, Shengju Wu2, Dean Ta1
1Department of Electronic Engineering, Fudan University, Shanghai 200433, China.
Abstract:
Ultrasonic guided wave (GW) assessment of long bone fracture have conventionally been based on pulse excitation. However, the high attenuation during propagation diminishes the amplitude of received GWs and results in low signal-to-noise ratio (SNR). The Barker code (BC) excitation and the optimal binary code (OBC) excitation were utilized in this study to overcome this limitation. Both simulations and in vitro experiments were performed on the fractured cortical bone plate model, and measured signals from both the BC and OBC excitations were decoded using the finite impulse response least squares inverse filter (FIR-LSIF) and then compared with sine pulse (SP) excited signals. The results suggest the efficiency of coded excitation for amplitude and SNR improvement. Furthermore, time-frequency representation (TFR) analysis was applied to experimental signals; with increasing fracture depth, energy transformation between predominate GW modes A1 and S2 was confirmed. These results show the potential of using BC and OBC excitations to evaluate the depth of long bone fracture.
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the...