Related Experiment Video
Updated: May 5, 2026

Method to Measure Tone of Axial and Proximal Muscle
Published on: December 14, 2011
The effect of rotating platform TKA on strain distribution and torque transmission on the proximal tibia
Robert A Malinzak1, Scott R Small1, Renee D Rogge2
1Joint Replacement Surgeons of Indiana Foundation, Inc., Mooresville, Indiana.
Abstract:
Limited experimental data exist comparing the mechanical response of the tibial cortex between fixed and rotating platform (RP) total knee arthroplasty (TKA), particularly in the revision setting. We asked if RP-TKA significantly affects tibiofemoral torque and cortical stain response in both the primary and revision settings. Fixed and RP tibial trays were implanted into analogue tibias and biomechanically tested under axial and torsional loading. Torque and strain response were analyzed using digital image correlation. Fixed bearing designs exhibited 13.8 times greater torque (P<0.01), and 69% (P<0.01) higher cortical strain than RP designs. Strain response was similar in the primary and revision cohorts. The decrease in torque transfer could act as a safeguard to reduce stress, micromotion and torsional fatigue in scenario of poor bone stock.
Related Concept Videos
Residual Stresses in Circular Shafts
Deformation in a Circular Shaft
Stress Concentrations in Circular Shafts
Normal Strain under Axial Loading
Thin-Walled Hollow Shafts
Design of Transmission Shafts - Stress Analysis