Estimation of background spectrum in a shielded HPGe detector using Monte Carlo simulations
1Experimental Nuclear Physics Department, Nuclear Research Centre, P.O. 13759, Cairo, Egypt.
Abstract:
Monte Carlo simulations are powerful tools used to estimate the background γ-radiation detected by high-resolution gamma-ray spectrometry systems with a HPGe (high purity germanium) detector contained inside a lead shield. The purpose of this work was to examine the applicability of Monte Carlo simulations to predict the optimal lead thickness necessary to reduce the background effect in spectrometer measurements. GEANT4 code was applied to simulate the background radiation spectrum at different thicknesses of lead. The simulated results were compared with experimental measurements of background radiation taken at the same shielding thickness. The results show that the background radiation detected depends on the thickness, size and lining of the shield. Simulation showed that 12 cm lead thick is the optimal shielding thickness.