Optimum AV interval in dual chamber pacemakers
Pacing and Clinical Electrophysiology : PACE
|September 1, 1986
Summary
Optimizing the atrioventricular (AV) interval in dual chamber pacemakers is crucial for maximizing cardiac output. The study found specific AV intervals (150-200 ms) yield the highest cardiac output in patients with complete heart block.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Devices
Background:
- Dual chamber pacemakers are widely used, with 50-70% of permanent implants being dual chamber.
- Optimal adjustment of the atrioventricular (AV) interval to maximize cardiac output in these patients is not well understood.
Purpose of the Study:
- To investigate the effect of varying atrioventricular (AV) intervals on cardiac output in patients with dual chamber pacemakers and complete heart block.
- To identify optimal AV interval settings for maximizing cardiac output at different pacing rates.
Main Methods:
- Ten patients with complete heart block and dual chamber pacemakers were studied.
- Cardiac output was measured using pulsed Doppler at pacing heart rates of 80, 100, and 118 beats/minute.
- Simultaneous measurements were taken across five different atrioventricular (AV) intervals.
Main Results:
- Maximum cardiac output was achieved with AV intervals of 150 ms and 200 ms at 80 and 100 beats/minute.
- An AV interval of 150 ms provided maximum cardiac output at 118 beats/minute.
- Increasing the AV interval to 250 ms significantly decreased cardiac output across all tested heart rates.
Conclusions:
- The study recommends noninvasive cardiac output measurement for personalized AV interval optimization in pacemaker patients.
- In the absence of direct measurement, an AV interval of 150 ms or 200 ms is suggested for achieving optimal cardiac output in most patients.