Related Experiment Video
Updated: May 5, 2026

A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
Screw fixation of medial malleolar fractures: a cadaveric biomechanical study challenging the current AO philosophy
L Parker1, N Garlick, I McCarthy
1The Royal Free Hospital, Pond Street, London NW3 2QG, UK.
Abstract:
The AO Foundation advocates the use of partially threaded lag screws in the fixation of fractures of the medial malleolus. However, their threads often bypass the radiodense physeal scar of the distal tibia, possibly failing to obtain more secure purchase and better compression of the fracture. We therefore hypothesised that the partially threaded screws commonly used to fix a medial malleolar fracture often provide suboptimal compression as a result of bypassing the physeal scar, and proposed that better compression of the fracture may be achieved with shorter partially threaded screws or fully threaded screws whose threads engage the physeal scar. We analysed compression at the fracture site in human cadaver medial malleoli treated with either 30 mm or 45 mm long partially threaded screws or 45 mm fully threaded screws. The median compression at the fracture site achieved with 30 mm partially threaded screws (0.95 kg/cm(2) (interquartile range (IQR) 0.8 to 1.2) and 45 mm fully threaded screws (1.0 kg/cm(2) (IQR 0.7 to 2.8)) was significantly higher than that achieved with 45 mm partially threaded screws (0.6 kg/cm(2) (IQR 0.2 to 0.9)) (p = 0.04 and p < 0.001, respectively). The fully threaded screws and the 30mm partially threaded screws were seen to engage the physeal scar under an image intensifier in each case. The results support the use of 30 mm partially threaded or 45 mm fully threaded screws that engage the physeal scar rather than longer partially threaded screws that do not. A 45 mm fully threaded screw may in practice offer additional benefit over 30 mm partially threaded screws in increasing the thread count in the denser paraphyseal region.
More Related Videos
07:41An Intramedullary Locking Nail for Standardized Fixation of Femur Osteotomies to Analyze Normal and Defective Bone Healing in Mice
Published on: November 13, 2016
04:41Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025