Related Experiment Video
Updated: May 5, 2026

A Recovery Cardiopulmonary Bypass Model Without Transfusion or Inotropic Agents in Rats
Published on: March 23, 2018
Emergent cardiopulmonary bypass for a 180 kilogram patient: support with a single oxygenator
Kyler T Hunter1, Angela R Lopez, Kwok L Yun
1Kaiser Permanente Los Angeles Medical Center, Regional Department of Cardiac Surgery, Los Angeles, California 90027, USA. kth2nd@gmail.com
Abstract:
As obesity increases in prevalence, so will cases in which patients present at the boundaries of care. We report the support of a class III obese man, having a body mass index of 60.8 kg/m2 and in acute renal failure, with a single Trillium-coated Affinity NT Hollow Fiber oxygenator in cardiopulmonary bypass for an emergent aortic valve replacement secondary to infective endocarditis. A maximum oxygen delivery of 807.51 mL of oxygen per minute is reported for this oxygenator in this case report.
More Related Videos
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Cardiomyopathy VII: Pre and Post Operative Nursing Management
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...