Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
The precision of ophthalmic biometry using calipers
Ashik Mohamed1, Derek Nankivil, Veerendranath Pesala
1Prof. Brien Holden Eye Research Centre, L V Prasad Eye Institute, Hyderabad, India; Vision Cooperative Research Centre, Brien Holden Vision Institute, Sydney, Australia; School of Optometry and Vision Science, University of New South Wales, Sydney, Australia.
This study found that calipers used in ophthalmic biometry have limited precision, around 0.1 mm for both globe and cornea measurements. Intraoperator and interoperator variance was consistent across different caliper types and measurement scales.
Area of Science:
- Ophthalmic biometry
- Ophthalmology
- Medical device precision
Background:
- Accurate ocular measurements are crucial for ophthalmic procedures.
- Calipers are commonly used instruments for biometry, but their precision requires evaluation.
Purpose of the Study:
- To determine the precision of whole globe and cornea measurements using calipers.
- To quantify the intraoperator and interoperator variance in these measurements.
Main Methods:
- Experimental study involving ten human donor eyes.
- Measurements of globe and cornea dimensions using digital Vernier and Castroviejo calipers.
- Three operators performed five repeat measurements for each dimension.
Main Results:
- No significant differences in globe measurements (anteroposterior length, diameters) were found.
- Horizontal corneal diameter consistently exceeded vertical diameter across all measurements.
- Variability was similar across operators and measurement scales, with no significant difference between caliper types.
- Mean intraoperator standard deviation was 0.127 ± 0.023 mm (digital caliper) and 0.094 ± 0.056 mm (Castroviejo caliper).
Conclusions:
- Commercially available calipers offer limited precision for ophthalmic biometry, approximately 0.1 mm.
- The precision is consistent across operators and measurement scales for both globe and cornea dimensions.
Related Concept Videos
Accuracy and Precision
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...