Removal of the antiscatter grid during routine biliary interventional procedures performed in a flat-panel
Kelvin Cortis1, Roberto Miraglia, Luigi Maruzzelli
1Diagnostic and Therapeutic Services, Mediterranean Institute for Transplantation and Advanced Specialized Therapies (ISMETT), Via Ernesto Tricomi 5, 90127, Palermo, Italy, kelvincortis@gmail.com.
Purpose:
To determine whether grid removal during routine biliary interventional procedures performed in a flat-panel interventional suite results in adequate image quality and a significant decrease of patient radiation exposure.
Materials And Methods:
Routine biliary interventional procedures were defined as those in which absence of fine image detail during fluoroscopy carries no procedural impact, including substitution of internal-external biliary drains (n = 25) or bilioplasty of benign biliary anastomotic strictures (n = 5). All patients had undergone a previous procedure in which the grid was used. Constant object-to-detector and source-to-image distance were maintained in each patient during the grid/no-grid procedures. The same fluoroscopy protocol was used for all examinations. The dose area product (DAP [cGy.cm(2)]) and procedure fluoroscopy time (seconds) were recorded for each procedure. DAP was normalized per unit of fluoroscopy time (nDAP [cGy.cm(2)/s]).
Results:
In all procedures, image quality was considered adequate by two different interventional radiologists, and all procedures were successfully completed without significant changes in fluoroscopy time between the two groups (p = 0.13). In every procedure without the grid, nDAP was inferior compared with nDAP in procedures performed using the grid. The mean decrease in dose was 39.2 ± 23.5 % (p = 0.000001).
Conclusion:
Our preliminary data show that removal of the grid during routine biliary procedures is feasible and results in a significant decrease of patient radiation exposure. This seems of particular relevance because most of these patients require frequent reintervention. Larger studies with more procedures are warranted to confirm these data.
More Related Videos
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
11:38Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
Related Concept Videos
Endoscopic Procedures V: ERCP
Patient...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...