Dose equations for tube current modulation in CT scanning and the interpretation of the associated CTDIvol
1Department of Radiology, Wake Forest University School of Medicine, Winston-Salem, North Carolina 27157.
Purpose:
The scanner-reported CTDI(vol) for automatic tube current modulation (TCM) has a different physical meaning from the traditional CTDI(vol) at constant mA, resulting in the dichotomy "CTDI(vol) of the first and second kinds" for which a physical interpretation is sought in hopes of establishing some commonality between the two.
Methods:
Rigorous equations are derived to describe the accumulated dose distributions for TCM. A comparison with formulae for scanner-reported CTDI(vol) clearly identifies the source of their differences. Graphical dose simulations are also provided for a variety of TCM tube current distributions (including constant mA), all having the same scanner-reported CTDI(vol).
Results:
These convolution equations and simulations show that the local dose at z depends only weakly on the local tube current i(z) due to the strong influence of scatter from all other locations along z, and that the "local CTDI(vol)(z)" does not represent a local dose but rather only a relative i(z) ≡ mA(z). TCM is a shift-variant technique to which the CTDI-paradigm does not apply and its application to TCM leads to a CTDI(vol) of the second kind which lacks relevance.
Conclusions:
While the traditional CTDI(vol) at constant mA conveys useful information (the peak dose at the center of the scan length), CTDI(vol) of the second kind conveys no useful information about the associated TCM dose distribution it purportedly represents and its physical interpretation remains elusive. On the other hand, the total energy absorbed E ("integral dose") as well as its surrogate DLP remain robust between variable i(z) TCM and constant current i0 techniques, both depending only on the total mAs = {i}t0 = i0t0 during the beam-on time t0.
More Related Videos
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Transmission-Line Differential Equations
Line Section Model
A circuit representing a line section of length Δx helps in understanding the transmission line parameters. The voltage V(x) and current i(x) are measured...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET