Initial results of using a novel irrigated multielectrode mapping and ablation catheter for pulmonary vein isolation
Dong-In Shin1, Kiriakos Kirmanoglou1, Christian Eickholt1
1Division of Cardiology, Pulmology and Vascular Medicine, University Hospital, Duesseldorf, Germany.
Background:
Pulmonary vein isolation (PVI) as a cornerstone for catheter ablation of atrial fibrillation (AF) remains a complex and time-consuming procedure.
Objective:
To assess feasibility, safety, and acute efficacy of a novel irrigated multielectrode ablation catheter guided by an electroanatomic mapping system for PVI in patients with paroxysmal AF.
Methods:
Twenty-five consecutive patients (60 ± 10 years) with paroxysmal AF underwent PVI by using a novel decapolar mapping and ablation catheter (nMARQ catheter, Biosense Webster Inc, Diamond Bar, CA). Ablation was guided by electroanatomic mapping allowing radiofrequency (RF) energy delivery in the antral region of pulmonary veins (PVs) from 10 irrigated electrodes simultaneously. The day after ablation, cardiac magnetic resonance imaging was performed routinely in order to assess for potential acute PV stenosis.
Results:
Overall, 97 of 97 (100%) targeted PVs could be isolated with a mean of 27 ± 11 RF applications and a mean total burning time of 15 ± 6 minutes per patient. The total procedure time from femoral vein access to catheter withdrawal was 110 ± 31 minutes, including a mean total fluoroscopy time of 23 ± 9 minutes. On average, 6 ± 3 RF impulses with a maximum of 25 W were applied per vein. After a short learning curve, procedure, fluoroscopy, and total burning times decreased to 94 ± 16, 16 ± 3, and 9 ± 2 minutes, respectively (P < .05). Entrance and exit blocks could be verified by placing the ablation catheter into 90 of 97 (93%) PVs in 18 of 25 (72%) patients. No procedure-related complications were observed, especially no acute PV stenosis could be detected.
Conclusions:
The use of a novel irrigated multielectrode ablation system for PVI is feasible and safe, resulting in acute isolation of all targeted PVs with no complications and short procedure times. Sustainability of these initial results has to be confirmed in long-term efficacy and follow-up trials.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
28:13Catheter Ablation in Combination With Left Atrial Appendage Closure for Atrial Fibrillation
Published on: February 26, 2013
Related Concept Videos
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization II: Right Heart Catheterization