Use of electrical pacing and automatic cardioversion defibrillation for normal cardiac function recovery
Pacing and Clinical Electrophysiology : PACE
|November 1, 1986
Summary
Antitachycardia pacing and low-energy cardioversion effectively treated ventricular arrhythmias in patients with heart conditions. An implantable cardioverter-defibrillator (ICAR-D) showed promise in preventing and terminating these dangerous heart rhythms.
Area of Science:
- Cardiology
- Biomedical Engineering
Background:
- Ventricular tachycardia (VT) and fibrillation (VF) are life-threatening arrhythmias.
- Acute myocardial infarction, coronary artery disease, and long QT syndrome increase arrhythmia risk.
- Current treatments include antitachycardia pacing and cardioversion.
Purpose of the Study:
- To evaluate the efficacy of antitachycardia pacing and low-energy cardioversion in terminating VT and VF.
- To assess the preventive effect on extrasystoles in high-risk patients.
- To investigate the effectiveness of a Soviet prototype implantable cardioverter-defibrillator (ICAR-D) for automatic arrhythmia detection and treatment.
Main Methods:
- Utilized antitachycardia pacing and low-energy cardioversion in patients.
- Employed low-energy automatic defibrillation in 25 patients intraoperatively and postoperatively.
- Tested the ICAR-D prototype in 10 dogs over 4-6 months in chronic experiments.
- ICAR-D delivers 2.5-4 J for VT and 15-20 J for VF.
Main Results:
- Antitachycardia pacing and low-energy cardioversion successfully terminated VT and VF episodes.
- These interventions also prevented extrasystoles in patients with specific cardiac conditions.
- The ICAR-D prototype demonstrated automatic arrhythmia detection and selective electrical impulse delivery.
Conclusions:
- Antitachycardia pacing and low-energy cardioversion are effective in managing VT and VF.
- The ICAR-D prototype shows potential as an automatic defibrillation device for preventing and treating cardiac arrhythmias.
- Further studies are warranted to confirm long-term efficacy and safety.