Fat-Water Phantoms for Magnetic Resonance Imaging Validation: A Flexible and Scalable Protocol
Published on: September 7, 2018
Relaxation effects in MRI-based quantification of fat content and fatty acid composition
Pernilla Peterson1, Jonas Svensson, Sven Månsson
1Department of Medical Radiation Physics, Malmö, Department of Clinical Sciences, Malmö, Lund University, Skåne University Hospital, Malmö, Sweden.
Purpose:
To investigate various sources of bias in MRI-based quantification of fat fraction (FF) and fatty acid composition (FAC) using chemical shift-encoded techniques.
Methods:
Signals from various FFs and FACs and individual relaxation rates of all signal components were simulated. From these signals, FF and FAC parameters were estimated with and without correction for differences in individual relaxation rates. In addition, phantom experiments were conducted with various flip angles and number of echoes to validate the simulations.
Results:
As expected, T(1) weighting resulted in an overestimation of the FF, but had much smaller impact on the FAC parameters. Differences in T(2) values of the signal components resulted in overestimation of the FAC parameters in fat/water mixtures, whereas the estimation in pure oil was largely unaffected. This bias was corrected using a simplified signal model with different T(2) values of water and fat, where the accuracy of the modeled T(2) of water was critical. The results of the phantom experiment were in agreement with simulations.
Conclusion:
T(1) weighting has only a minor effect on FAC quantification in both fat/water mixtures and pure oils. T(2) weighting is mainly a concern in fat/water mixtures but may be corrected using a simplified model.