Use of non-invasive total hemoglobin measurement as a screening tool for anemia in children

Izuki Amano1, Ayako Murakami

  • 1Akitsu Children's Clinic, Tokyo, Japan.

Insights

Non-invasive hemoglobin (SpHb) screening using the Radical-7® Pulse CO-Oximeter™ shows promise for pediatric anemia detection. While further refinement is needed, this method offers a less invasive alternative to traditional blood sampling for children.

Area of Science:

  • Pediatric Medicine
  • Medical Devices
  • Diagnostic Technology

Background:

  • Universal anemia screening in children is crucial but often requires invasive blood sampling.
  • Non-invasive hemoglobin measurement devices aim to simplify this process.
  • The clinical utility of the Radical-7® Pulse CO-Oximeter™ for pediatric anemia screening remains to be fully established.

Purpose of the Study:

  • To evaluate the effectiveness of the Radical-7® Pulse CO-Oximeter™ for non-invasive hemoglobin concentration (SpHb) measurement in children.
  • To compare SpHb readings with results from a standard complete blood count (CBC) analysis.
  • To assess the feasibility of using SpHb as a primary screening tool for anemia in pediatric populations.

Main Methods:

  • SpHb measurements were performed using the Radical-7® Pulse CO-Oximeter™ during routine medical check-ups for 110 three-year-old children.
  • A separate cohort of 43 pediatric patients underwent both CBC analysis using a Microsemi® hematology analyzer and Radical-7® monitoring.
  • Statistical analysis included correlation and Bland-Altman comparison to assess agreement between SpHb and CBC.

Main Results:

  • The mean SpHb level in the 3-year-old cohort was 12.1 ± 0.64 g/dL.
  • A significant correlation (r=0.602, P < 0.0001) was observed between Radical-7® SpHb and Microsemi® CBC results.
  • Bland-Altman analysis indicated a bias of -0.6 ± 1.1 g/dL between the two measurement methods.

Conclusions:

  • The Radical-7® Pulse CO-Oximeter™ demonstrates potential as a non-invasive tool for initial anemia screening in children.
  • While the device shows promise, further improvements are necessary to enhance accuracy and reliability.
  • Non-invasive SpHb measurement offers a viable alternative for primary screening, potentially reducing the need for invasive procedures.