Use of non-invasive total hemoglobin measurement as a screening tool for anemia in children
1Akitsu Children's Clinic, Tokyo, Japan.
Insights
Non-invasive hemoglobin (SpHb) screening using the Radical-7® Pulse CO-Oximeter™ shows promise for pediatric anemia detection. While further refinement is needed, this method offers a less invasive alternative to traditional blood sampling for children.
Area of Science:
- Pediatric Medicine
- Medical Devices
- Diagnostic Technology
Background:
- Universal anemia screening in children is crucial but often requires invasive blood sampling.
- Non-invasive hemoglobin measurement devices aim to simplify this process.
- The clinical utility of the Radical-7® Pulse CO-Oximeter™ for pediatric anemia screening remains to be fully established.
Purpose of the Study:
- To evaluate the effectiveness of the Radical-7® Pulse CO-Oximeter™ for non-invasive hemoglobin concentration (SpHb) measurement in children.
- To compare SpHb readings with results from a standard complete blood count (CBC) analysis.
- To assess the feasibility of using SpHb as a primary screening tool for anemia in pediatric populations.
Main Methods:
- SpHb measurements were performed using the Radical-7® Pulse CO-Oximeter™ during routine medical check-ups for 110 three-year-old children.
- A separate cohort of 43 pediatric patients underwent both CBC analysis using a Microsemi® hematology analyzer and Radical-7® monitoring.
- Statistical analysis included correlation and Bland-Altman comparison to assess agreement between SpHb and CBC.
Main Results:
- The mean SpHb level in the 3-year-old cohort was 12.1 ± 0.64 g/dL.
- A significant correlation (r=0.602, P < 0.0001) was observed between Radical-7® SpHb and Microsemi® CBC results.
- Bland-Altman analysis indicated a bias of -0.6 ± 1.1 g/dL between the two measurement methods.
Conclusions:
- The Radical-7® Pulse CO-Oximeter™ demonstrates potential as a non-invasive tool for initial anemia screening in children.
- While the device shows promise, further improvements are necessary to enhance accuracy and reliability.
- Non-invasive SpHb measurement offers a viable alternative for primary screening, potentially reducing the need for invasive procedures.
Abstract:
Universal screening for anemia is important in children, but invasive blood sampling is required. A new device (Radical-7® Pulse CO-Oximeter™, Masimo, Irvine, CA, USA) now enables non-invasive hemoglobin concentration (SpHb) measurement to be done, but the usefulness of this device for anemia screening in children is unclear. The objective of this study was to compare SpHb with complete blood count (CBC) using a hematology analyzer (Microsemi® LC-667CRP; Fukuda Denshi, Tokyo, Japan). SpHb measurement with Radical-7® was done as part of a medical check-up in 3-year-old children (n = 110). Another 43 pediatric patients were checked for CBC using Microsemi® and monitored with Radical-7®. The mean SpHb level of the 3-year-old children was 12.1 ± 0.64 g/dL (range, 10.8-13.7 g/dL). The correlation of Radical-7® and Microsemi® was 0.602 (P < 0.0001). On Bland-Altman comparison, bias was -0.6 ± 1.1 g/dL. Even though further improvement is required, Radical-7® offers many possibilities in the context of primary screening.