Phantom and clinical study of differences in cone beam computed tomographic registration when aligned to maximum and
Kiyonori Shirai1, Kinji Nishiyama1, Toshizo Katsuda2
1Department of Radiation Oncology, Osaka Medical Center for Cancer and Cardiovascular Diseases, Osaka, Japan.
Purpose:
To determine whether maximum or average intensity projection (MIP or AIP, respectively) reconstructed from 4-dimensional computed tomography (4DCT) is preferred for alignment to cone beam CT (CBCT) images in lung stereotactic body radiation therapy.
Methods And Materials:
Stationary CT and 4DCT images were acquired with a target phantom at the center of motion and moving along the superior-inferior (SI) direction, respectively. Motion profiles were asymmetrical waveforms with amplitudes of 10, 15, and 20 mm and a 4-second cycle. Stationary CBCT and dynamic CBCT images were acquired in the same manner as stationary CT and 4DCT images. Stationary CBCT was aligned to stationary CT, and the couch position was used as the baseline. Dynamic CBCT was aligned to the MIP and AIP of corresponding amplitudes. Registration error was defined as the SI deviation of the couch position from the baseline. In 16 patients with isolated lung lesions, free-breathing CBCT (FBCBCT) was registered to AIP and MIP (64 sessions in total), and the difference in couch shifts was calculated.
Results:
In the phantom study, registration errors were within 0.1 mm for AIP and 1.5 to 1.8 mm toward the inferior direction for MIP. In the patient study, the difference in the couch shifts (mean, range) was insignificant in the right-left (0.0 mm, ≤1.0 mm) and anterior-posterior (0.0 mm, ≤2.1 mm) directions. In the SI direction, however, the couch position significantly shifted in the inferior direction after MIP registration compared with after AIP registration (mean, -0.6 mm; ranging 1.7 mm to the superior side and 3.5 mm to the inferior side, P=.02).
Conclusions:
AIP is recommended as the reference image for registration to FBCBCT when target alignment is performed in the presence of asymmetrical respiratory motion, whereas MIP causes systematic target positioning error.
More Related Videos
07:13Author Spotlight: An Efficient and Robust Software for Automated Fusion of Multiple Preclinical Imaging Modalities
Published on: October 27, 2023
10:23Author Spotlight: Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
Published on: September 8, 2023
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET