Gout tophus detection-a comparison of dual-energy CT (DECT) and histology
Ralph Melzer1, Chantal Pauli2, Thomas Treumann3
1Clinic of Rheumatology, Luzerner Kantonsspital, 6000 Lucerne 16, Switzerland.
Objectives:
To compare dual-energy CT images (DECT) directly with gold standard histology.
Methods:
A 85-year-old woman with chronic recurrent gouty arthritis and macroscopically visible tophi was assessed with DECT-scans of her feet. When she died 7 days later, three tophi in different regions and in different tissues of her left foot were processed for histological examination.
Results:
Of the selected tophi, two were almost completely missed by the color-coding DECT sequences. The tophi remained in most of their volume below the detection threshold value of 150HU (default value).
Conclusions:
It could be demonstrated that DECT only highlights the dense tophi (corresponding of approximately 15-20vol% urate in the tophus). Less dense tophi, despite considerable size, will be missed in the color-coded images.
More Related Videos
07:013D Imaging of Soft-Tissue Samples using an X-ray Specific Staining Method and Nanoscopic Computed Tomography
Published on: October 24, 2019
10:28Gene Regulation and Targeted Therapy in Gastric Cancer Peritoneal Metastasis: Radiological Findings from Dual Energy CT and PET/CT
Published on: January 22, 2018
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Imaging Studies III: Computed Tomography
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...