Axial biometry of the entire eye using ultra-long scan depth optical coherence tomography
Jianguang Zhong1, Yilei Shao2, Aizhu Tao2
1Bascom Palmer Eye Institute, University of Miami Miller School of Medicine, Miami, Florida; Department of Ophthalmology, Hangzhou First People's Hospital, Hangzhou, Zhejiang, China.
Purpose:
To assess the repeatability of axial biometry of the entire eye using ultra-long scan depth optical coherence tomography (OCT) and to investigate the agreement with IOLMaster measurements (Carl Zeiss Meditec).
Design:
Prospective, observational case series.
Methods:
There were 37 adult subjects enrolled in group 1 and 12 adult subjects enrolled in group 2. Using ultra-long scan depth OCT, the left eyes of these groups were measured in 2 separate sessions. The images were processed by a manual method and custom-developed automatic software. A model eye was imaged for verification. The subjects in group 2 were imaged using ultra-long scan depth OCT and using the IOLMaster for axial length measurement comparison.
Results:
All measured parameters of the model eye matched the geometric parameters. In group 1, there were no significant differences in all measured parameters using automatic and manual segmentation methods (P > .05, paired t test). The percentage of coefficient of repeatability of segments ranged from 0.3% to 3.9%. The corresponding interclass correlation coefficients ranged from 0.946 to 0.999. The correlation between the results using automatic and manual segmentation methods appeared to be strong (R(2) = 0.999; P < .05). In group 2, the axial length of the eye measured by the IOLMaster matched the results obtained by ultra-long scan depth OCT with the automatic method (R1(2) = 0.987; P < .05) and the manual method (R2(2) = 0.988; P < .05).
Conclusions:
Automatic axial biometry using ultra-long scan depth OCT successfully measured each segment of the entire eye with good repeatability. With further development of automatic segmentation, ultra-long scan depth OCT seems to be a promising tool in the axial biometry of the entire eye.
More Related Videos
10:10Ultrahigh Resolution Mouse Optical Coherence Tomography to Aid Intraocular Injection in Retinal Gene Therapy Research
Published on: November 2, 2018
12:22Multimodal Volumetric Retinal Imaging by Oblique Scanning Laser Ophthalmoscopy oSLO and Optical Coherence Tomography OCT
Published on: August 4, 2018
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...