Related Experiment Video
Updated: May 4, 2026

Graphene Coatings for Biomedical Implants
Published on: March 1, 2013
Graphene oxide doped conducting polymer nanocomposite film for electrode-tissue interface
Hong-Chang Tian1, Jing-Quan Liu1, Dai-Xu Wei2
1National Key Laboratory of Science and Technology on Micro/Nano Fabrication, Shanghai Jiao Tong University, Shanghai, PR China.
Abstract:
One of the most significant components for implantable bioelectronic devices is the interface between the microelectrodes and the tissue or cells for disease diagnosis or treatment. To make the devices work efficiently and safely in vivo, the electrode-tissue interface should not only be confined in micro scale, but also possesses excellent electrochemical characteristic, stability and biocompatibility. Considering the enhancement of many composite materials by combining graphene oxide (GO) for its multiple advantages, we dope graphene oxide into poly(3,4-ethylenedioxythiophene) (PEDOT) forming a composite film by electrochemical deposition for electrode site modification. As a consequence, not only the enlargement of efficient surface area, but also the development of impedance, charge storage capacity and charge injection limit contribute to the excellent electrochemical performance. Furthermore, the stability and biocompatibility are confirmed by numerously repeated usage test and cell proliferation and attachment examination, respectively. As electrode-tissue interface, this biomaterial opens a new gate for tissue engineering and implantable electrophysiological devices.

