Curtain Flow Column: Optimization of Efficiency and Sensitivity
Published on: June 12, 2016
Detailed characterization of the kinetic performance of first and second generation silica monolithic columns for
Deirdre Cabooter1, Ken Broeckhoven2, Roman Sterken1
1KU Leuven, Department of Pharmaceutical Sciences, Pharmaceutical Analysis, Herestraat 49, 3000 Leuven, Belgium.
Abstract:
The kinetic performance of commercially available first generation and prototype second generation silica monoliths has been investigated for 2.0mm and 3.0-3.2mm inner diameter columns. It is demonstrated that the altered sol-gel process employed for the production of second generation monoliths results in structures with a smaller characteristic size leading to an improved peak shape and higher efficiencies. The permeability of the columns however, decreases significantly due to the smaller throughpore and skeleton sizes. Scanning electron microscopy pictures suggest the first generation monoliths have cylindrical skeleton branches, whereas the second generation monoliths rather have skeleton branches that resemble a single chain of spherical globules. Using recently established correlations for the flow resistance of cylindrical and globule chain type monolithic structures, it is demonstrated that the higher flow resistance of the second generation monoliths can be entirely attributed to their smaller skeleton sizes, which is also evident from the external porosity that is largely the same for both monolith generations (ɛe∼0.65). The recorded van Deemter plots show a clear improvement in efficiency for the second generation monoliths (minimal plate heights of 13.6-14.1μm for the first and 6.5-8.2μm for the second generation, when assessing the plate count using the Foley-Dorsey method). The corresponding kinetic plots, however, indicate that the much reduced permeability of the second generation monoliths results in kinetic performances (time needed to achieve a given efficiency) which are only better than those of the first generation for plate counts up to N∼45,000. For more complex samples (N≥50,000), the first generation monoliths can intrinsically still provide faster analysis due to their high permeability. It is also demonstrated that - despite the improved efficiency of the second generation monoliths in the practical range of separations (N=10,000-50,000) - these columns can still not compete with state-of-the-art core-shell particle columns when all columns are evaluated at their own maximum operating pressure (200bar for the monolithic columns, 600bar for core-shell columns). It is suggested that monolithic columns will only become competitive with these high efficiency particle columns when further improvements to their production process are made and their pressure resistance is raised.
Related Concept Videos
Silica Gel Column Chromatography: Overview
Polar components tend to bind strongly to the silica gel, causing them to move slowly through the column. In contrast, nonpolar compounds...
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Principles Of Column Chromatography
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Ion-Exchange Chromatography