Comparison of 3D cube FLAIR with 2D FLAIR for multiple sclerosis imaging at 3 Tesla
M Patzig1, M Burke2, H Brückmann1
1Dept. of Neuroradiology, University of Munich.
Purpose:
Three-dimensional (3 D) MRI sequences allow improved spatial resolution with good signal and contrast properties as well as multiplanar reconstruction. We sought to compare Cube, a 3 D FLAIR sequence, to a standard 2 D FLAIR sequence in multiple sclerosis (MS) imaging.
Materials And Methods:
Examinations were performed in the clinical routine on a 3.0 Tesla scanner. 12 patients with definite MS were included. Lesions with MS-typical properties on the images of Cube FLAIR and 2 D FLAIR sequences were counted and allocated to different brain regions. Signal-to-noise ratios (SNR) and contrast-to-noise ratios (CNR) were calculated.
Results:
With 384 the overall number of lesions found with Cube FLAIR was significantly higher than with 2 D FLAIR (N = 221). The difference was mostly accounted for by supratentorial lesions (N = 372 vs. N = 216) while the infratentorial lesion counts were low in both sequences. SNRs and CNRs were significantly higher in CUBE FLAIR with the exception of the CNR of lesion to gray matter, which was not significantly different.
Conclusion:
Cube FLAIR showed a higher sensitivity for MS lesions compared to a 2 D FLAIR sequence. 3 D FLAIR might replace 2 D FLAIR sequences in MS imaging in the future.
More Related Videos
Related Concept Videos
Imaging Studies III: Computed Tomography
Magnetic Resonance Imaging
Imaging Studies IV: Magnetic Resonance Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies for Cardiovascular System IV: CMRI