Implantable cardioverter defibrillator and inappropriate therapy: "black box" examination yielded both human and
1Electrophysiology Unit, Cardiology division, USEK- NDS University Hospital, St Charbel Street, Byblos, Lebanon.
Abstract:
We report on a 70-year-old male patient who was recipient of GEM III DR 7275 Cardioverter Defibrillator, and who presented with inappropriate shocks. The patient had a documented slow ventricular tachycardia (VT), and the device was programmed to detect VT at rates >100 bpm, fast VT (FVT, via VT) at rates >150 bpm, and ventricular fibrillation (VF) at rates >188 bpm. After detection of FVT, efficient therapy was delivered; however, this was immediately followed by multiple inappropriate therapies. Inappropriate therapies were discussed, with a focus on programming features.