Related Experiment Video
Updated: May 4, 2026

Quantitative and Temporal Control of Oxygen Microenvironment at the Single Islet Level
Published on: November 17, 2013
Periodic-zone model predictive control for diurnal closed-loop operation of an artificial pancreas
Ravi Gondhalekar1, Eyal Dassau, Howard C Zisser
1University of California, Santa Barbara, Santa Barbara, California; and Sansum Diabetes Research Institute, Santa Barbara, California.
Background:
The objective of this research is an artificial pancreas (AP) that performs automatic regulation of blood glucose levels in people with type 1 diabetes mellitus. This article describes a control strategy that performs algorithmic insulin dosing for maintaining safe blood glucose levels over prolonged, overnight periods of time and furthermore was designed with outpatient, multiday deployment in mind. Of particular concern is the prevention of nocturnal hypoglycemia, because during sleep, subjects cannot monitor themselves and may not respond to alarms. An AP intended for prolonged and unsupervised outpatient deployment must strategically reduce the risk of hypoglycemia during times of sleep, without requiring user interaction.
Methods:
A diurnal insulin delivery strategy based on predictive control methods is proposed. The so-called "periodic-zone model predictive control" (PZMPC) strategy employs periodically time-dependent blood glucose output target zones and furthermore enforces periodically time-dependent insulin input constraints to modulate its behavior based on the time of day.
Results:
The proposed strategy was evaluated through an extensive simulation-based study and a preliminary clinical trial. Results indicate that the proposed method delivers insulin more conservatively during nighttime than during daytime while maintaining safe blood glucose levels at all times. In clinical trials, the proposed strategy delivered 77% of the amount of insulin delivered by a time-invariant control strategy; specifically, it delivered on average 1.23 U below, compared with 0.31 U above, the nominal basal rate overnight while maintaining comparable, and safe, blood glucose values.
Conclusions:
The proposed PZMPC algorithm strategically prevents nocturnal hypoglycemia and is considered a significant step toward deploying APs into outpatient environments for extended periods of time in full closed-loop operation.
Related Concept Videos
Time and frequency -Domain Interpretation of PI Control
Acting as a low-pass filter, the PI controller slows the system's response and extends settling times. This requires...
PI Controller: Design
Open and closed-loop control systems
An open-loop control system operates without feedback from the output. It consists of two primary elements: the controller and the controlled process. The controller receives an input signal...
PD Controller: Design
Designing a continuous-data controller requires selecting and linking components like adders and integrators, which are fundamental in Proportional,...
Glucose Homeostasis: Pancreatic Islets and Insulin Secretion
Insulin and C-peptide are...