Manufacturer changes lead to clinically important differences between two editions of the TNO stereotest
Louise L A van Doorn1, Bruce J W Evans, David F Edgar
1Damme Optometrie, Kesteren, The Netherlands; Department of Optometry, Hogeschool Utrecht, University of Applied Science, Utrecht, The Netherlands; City University London, London, UK.
Purpose:
Stereoacuity tests used in clinical practice should be repeatable and reproducible. However, it has been observed in a clinical setting that new editions of the TNO stereotest appear to give different values from those obtained using previous versions. The purpose of the present research was to investigate this observation.
Methods:
One hundred and twenty-one Dutch subjects, 88 (73%) females and 33 (27%) males, with an average age of 34.0 years (range 18-55) had their stereoacuity measured using two different versions of the TNO stereoacuity test (TNO 13 and TNO 15). The TNO was tested in a counterbalanced order so that consecutive subjects started with alternate editions to avoid bias.
Results:
There was a significant difference (p < 0.001) between the median value for stereoacuity measured with TNO 13 (30 s of arc) and TNO 15 (60 s of arc). The bias between the two test versions was -0.23 Log arcseconds (95% limits of the differences: 0.15 to -0.60 Log arcseconds).
Conclusion:
This study reveals that results obtained with two different editions of a commonplace stereoacuity test are not comparable. New versions come on the market at regular intervals and the assumption that they will give the same results as previous versions may not be valid. Besides the statistically significant difference between the TNO 13 and TNO 15, the Bland-Altman plot also showed a considerable bias and the 95% limits of the differences between the TNO 13 and TNO 15 are more than two steps on the Log arcsecond scale. This difference between two editions of the TNO stereotests is not clinically acceptable and therefore it is inappropriate to use the two versions of the test interchangeably. It is important in both research and clinical records to specify the edition of the TNO test used.
More Related Videos
05:57Novel Object Recognition and Object Location Behavioral Testing in Mice on a Budget
Published on: November 20, 2018
07:02A Computerized Test Battery to Study Pharmacodynamic Effects on the Central Nervous System of Cholinergic Drugs in Early Phase Drug Development
Published on: February 11, 2019
Related Concept Videos
Comparing Experimental Results: Student's t-Test
Pharmaceutical Alternatives: Stability-Related Therapeutic Nonequivalence
Statistical Methods to Analyze Parametric Data: Student t-Test and Goodness-of-Fit Test
The Student's t-test is a statistical test that examines if there is a statistically significant difference between the means of two groups. This test is instrumental when dealing with...
Testing a Claim about Standard Deviation
The hypothesis testing for the claim of population standard deviation (or variance) requires the data and samples to be random and unbiased. The population distribution also must be normal. There is no specific requirement on the sample size as the estimation is based on the chi-square distribution.
As a first step, the hypothesis (null and alternative) concerning the claim about...
Significance Testing: Overview
Behrens–Fisher Test
This test...