Related Experiment Video
Updated: May 4, 2026

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
A simple method for determining the activity of large-area beta sources constructed from anodized aluminum foils
1National Institute of R&D for Physics and Nuclear Engineering "HoriaHulubei", IFIN-HH, P.O.Box MG-6, Bucharest-Magurele R-077125, Romania.
Abstract:
A simple method has been developed for determining the activity of large-area beta reference sources in anodized aluminum foils. It is based on the modeling of the transmission of beta rays through thin foils in planar geometry using Monte Carlo simulation. The method was checked experimentally and measurement results show that the activity of large-area beta reference sources in anodized aluminum foils can be measured with standard uncertainties smaller than the limit of 10% required by ISO 8769.