Related Experiment Video
Updated: May 4, 2026

Ultrasensitive Detection of Biomarkers by Using a Molecular Imprinting Based Capacitive Biosensor
Published on: February 16, 2018
Contactless conductivity detection for screening myrosinase substrates by capillary electrophoresis
Reine Nehmé1, Hala Nehmé1, Grégory Roux1
1Institut de Chimie Organique et Analytique (ICOA), Université d'Orléans, CNRS FR 2708, UMR 7311, Orléans, France.
Abstract:
Myrosinase is a unique enzyme that catalyzes the hydrolysis of glucosinolates (GLS) to isothiocyanate (ITC), glucose and sulfate. Isothiocyanates display a diversified very interesting biological activity. In this study, capillary electrophoresis (CE) was used for the first time for evaluating myrosinase kinetics (maximum velocity Vmax and Michaelis-Menten constant Km) and to assess the affinity of a variety of substrates toward this enzyme. The pre-capillary approach was chosen since it is very simple to conduct. For this, the enzymatic reaction was performed in a micro-vial. The reaction mixture volume was of only 100 μL and the incubation lasted only 5 min at 37±1°C. Short-end injection of few tens of nanoliters (~25 nL) of the reaction mixture was performed which decreased analysis time without using any electroosmotic modifier. The sulfate produced was detected and quantified with a contactless capacitively coupled conductivity detector (C(4)D) allowing the evaluation of myrosinase kinetics. This study shows, that capillary electrophoresis with contactless conductivity detection can be very useful for monitoring myrosinase activity. Comparing to the conventional spectrophotometric method (1982), the CE method developed here is simple, automated, economic, rapid (incubation for few minutes) and robust. Results compared very well with those reported in literature using the conventional method. Moreover, the affinity of a variety of natural and synthetic glucosinolates toward this enzyme has been assessed for the first time.
More Related Videos
13:42Dry Film Photoresist-based Electrochemical Microfluidic Biosensor Platform: Device Fabrication, On-chip Assay Preparation, and System Operation
Published on: September 19, 2017
12:11ELIME Enzyme Linked Immuno Magnetic Electrochemical Method for Mycotoxin Detection
Published on: October 23, 2009
Related Concept Videos
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Capillary Electrophoresis: Instrumentation
Electrophoresis: Overview
There...