Background:
Continuous renal replacement therapy (CRRT) is an important tool in the care of critically ill patients. However, the impact of a specific CRRT machine type on the successful delivery of CRRT is unclear. The purpose of this study was to evaluate the effectiveness of CRRT delivery with an intensive care unit (ICU) bedside nurse delivery model for CRRT while comparing circuit patency and circuit exchange rates in 2 Food and Drug Administration-approved CRRT devices. This article presents the data comparing circuit exchange rates for 2 different CRRT machines.
Materials And Methods:
A group of ICU nurses were selected to undergo expanded training in CRRT operation and empowered to deliver all aspects of CRRT. The ICU nurses then provided all aspects of CRRT on 2 Food and Drug Administration-approved CRRT devices for 6 months. Each device was used exclusively in the designated ICU for a 2-week run-in period followed by 3-month data collection period. The primary end point for the study was the differences in average number of filter exchanges per day during each CRRT event.
Results:
A total of 45 unique patients who underwent 64 separate CRRT treatment periods were included. Four CRRT events were excluded (see text for details). Twenty-eight CRRT events occurred in the NxStage System One arm (NxStage Medical, Lawrence, Mass) and 32 events in the Gambro Prismaflex arm (Gambro Renal Products, Boulder, Colo). Average (SD) filter exchanges per day was 0.443 (0.60) for the NxStage System One machine and 0.553 (0.65) for Gambro Prismaflex machine (P = .09).
Conclusions:
There was no demonstrable difference in circuit patency as defined by the rate of filter exchanges per day of CRRT therapy.
Related Concept Videos
Extracorporeal Removal of Drugs: Continuous Renal Replacement Therapy
Continuous Renal Replacement Therapy
Peritoneal Dialysis II: Peritoneal Dialysis Systems and Complications
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...
Drug Dosing in Renal Diseases: Dose Adjustments Based on Drug Clearance and Elimination Rate Constant
Extracorporeal Removal of Drugs: Peritoneal Dialysis and Hemodialysis