A comparison of oxygen saturation data in inpatients with low oxygen saturation using automated continuous monitoring
Andreas H Taenzer1, Joshua Pyke, Michael D Herrick
1From the Department of Anesthesiology and Pediatrics, The Geisel School of Medicine at Dartmouth, Dartmouth Hitchcock Medical Center, Lebanon; Thayer School of Engineering, Dartmouth, Hanover; and Department of Anesthesiology, The Geisel School of Medicine at Dartmouth, Dartmouth Hitchcock Medical Center, Lebanon, New Hampshire.
Background:
The manual collection and charting of traditional vital signs data in inpatient populations have been shown to be inaccurate when compared with true physiologic values. This issue has not been examined with respect to oxygen saturation data despite the increased use of this measurement in systems designed to assess the risk of patient deterioration. Of particular note are the lack of available data examining the accuracy of oxygen saturation charting in a particularly vulnerable group of patients who have prolonged oxygen desaturations (mean SpO2 <90% over at least 15 minutes). In addition, no data are currently available that investigate the often suspected "wake up" effect, resulting from a nurse entering a patient's room to obtain vital signs.
Methods:
In this study, we compared oxygen saturation data recorded manually with data collected by an automated continuous monitoring system in 16 inpatients considered to be at high risk for deterioration (average SpO2 values <90% collected by the automated system in a 15-minute interval before a manual charting event). Data were sampled from the automatic collection system from 2 periods: over a 15-minute period that ended 5 minutes before the time of the manual data collection and charting, and over a 5-minute range before and after the time of the manual data collection and charting. Average saturations from prolonged baseline desaturations (15-minute period) were compared with both the manual and automated data sampled at the time of the nurse's visit to analyze for systematic change and to investigate the presence of an arousal effect.
Results:
The manually charted data were higher than those recorded by the automated system. Manually recorded data were on average 6.5% (confidence interval, 4.0%-9.0%) higher in oxygen saturation. No significant arousal effect resulting from the nurse's visit to the patient's room was detected.
Conclusions:
In a cohort of patients with prolonged desaturations, manual recordings of SpO2 did not reflect physiologic patient state when compared with continuous automated sampling. Currently, early warning scores depend on manual vital sign recordings in many settings; the study data suggest that SpO2 ought to be added to the list of vital sign values that have been shown to be recorded inaccurately.
More Related Videos
06:57Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...