Daniel F Fisher1, Christopher T Chenelle2, Andrew D Marchese3
1Respiratory Care Services, Massachusetts General Hospital, Boston, Massachusetts. dfisher2@partners.org.
Background:
Tracheal intubation is used to establish a secure airway in patients who require mechanical ventilation. Unexpected extubation can have serious complications, including airway trauma and death. Various methods and devices have been developed to maintain endotracheal tube (ETT) security. Associated complications include pressure ulcers due to decreased tissue perfusion. Device consideration includes ease of use, rapid application, and low exerted pressure around the airway.
Methods:
Sixteen ETT holders were evaluated under a series of simulated clinical conditions. ETT security was tested by measuring distance displaced after a tug. Nine of the 16 methods could be evaluated for speed of moving the ETT to the opposite side of the mouth. Sensors located on a mannequin measured applied forces when the head was rotated vertically or horizontally. Data were analyzed using multivariate analysis of variance, with P < .05.
Results:
Median displacement of the ETT by the tug test was 0 cm (interquartile range of 0.0-0.10 cm, P < .001). The mean time to move the ETT from one side of the mouth to the other ranged from 1.25 ± 0.2 s to 34.4 ± 3.4 s (P < .001). Forces applied to the face with a vertical head lift ranged from < 0.2 newtons (N) to a maximum of 3.52 N (P < .001). Forces applied to the face with a horizontal rotation ranged from < 0.2 N to 3.52 N (P < .001). Commercial devices produced greater force than noncommercial devices.
Conclusions:
Noncommercial airway holders exert less force on a patient's face than commercial devices. Airway stability is affected by the type of securing method. Many commercial holders allow for rapid but secure movement of the artificial airway from one side of the mouth to the other.
More Related Videos
04:43Endotracheal Intubation via Tracheotomy and Subsequent Thoracotomy in Rats for Non-Survival Applications
Published on: March 15, 2024
04:30Endotracheal Intubation Using a Flexible Intubation Endoscope as a Standardized Model for Safe Airway Management in Swine
Published on: August 25, 2022
Related Concept Videos
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Tracheostomy Suctioning II: Procedure
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...