Effect of solution properties on the counting and sizing of subvisible particle standards as measured by light
Tobias Werk1, David B Volkin2, Hanns-Christian Mahler1
1Pharmaceutical Development & Supplies, Pharma Technical Development Biologics EU, F. Hoffmann-La Roche Ltd., Basel, Switzerland.
Purpose:
Protein formulations may contain subvisible particle (SbVP) impurities that can vary (e.g., in number, size, shape, density, refractive index and transparency) depending on the formulation composition, environmental stresses and the type of protein. Additionally formulation solutions may differ in their physical properties including turbidity, color, viscosity, density and refractive index. This study examined the impact of these formulation matrix parameters on the ability to size and count subvisible particles using a variety of analytical methods including two light obscuration (HIAC, Syringe) and two digital imaging instruments (MFI®, FlowCAM®). Several subvisible particle standards were tested, including polystyrene and glass beads as well as a new pseudo-protein particle standard, in order to also study of the effect of subvisible particles with different properties.
Results:
The color and turbidity of solutions generally had a relatively small effect on SbVP sizing and counting. Solution viscosity and refractive index (RI), however, showed a more pronounced effect on the analytical results, especially with more translucent particles such as glass beads and the "pseudo protein standards", resulting in smaller sizes and lower counts of SbVPs, especially when measuring particles using light obscuration methods.
Conclusions:
Each instrument showed certain advantages and disadvantages depending on the analytical parameter (i.e., accuracy, precision), type of subvisible particle, and solution properties. Based on these results, it is recommended to not only carefully consider physical solution parameters as part of analytical method assessment for counting and sizing SbVP in protein dosage forms, but also in terms of various typical QC validation parameters using actual protein formulations.
Related Concept Videos
Microbial Growth Measurement: Indirect Methods
Precipitate Formation and Particle Size Control
The obtained precipitate should be either a pure substance of known composition or easily converted to one by a simple process, such as ignition or drying. In addition, the precipitate should be insoluble and easily filterable. In general, filterability...
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Microbial Growth Measurement: Direct Methods