Aim:
Three-dimensional high-resolution anorectal manometry (3DHRAM) is a new technique that can simultaneously provide physiological and topographical data on the terminal part of the digestive tract. Our object was to assess whether 3DHRAM is able to reliably diagnose excessive perineal descent already diagnosed with conventional defaecography, which is considered to be the gold standard.
Method:
All patients referred to our centre for anorectal manometry and conventional defaecography were evaluated with a maximum of 6 months between the two examinations. Anorectal manometry was performed using the 3D High-Resolution Given Imaging® probe. Excessive perineal descent was defined as the downward movement of the anal high-pressure zone during straining. At the end of the straining effort, the high-pressure zone regained its initial position, thereby indicating that the probe had not moved.
Results:
Nineteen female patients of median age 53 (21-70) years were included in the study. All cases with excessive perineal descent diagnosed using defaecography were visualized with 3DHRAM. The degree of perineal descent determined by 3D and conventional defaecography was compared (Spearman correlation 0.726, P = 0.01). In contrast, the averages measured were significantly different; the average was 11.68 ± 3.3 mm for 3DHRAM but 34.21 ± 13.3 mm for conventional defaecography (P = 0.002).
Conclusion:
The results of the study demonstrate that 3DHRAM can diagnose excessive perineal descent with the same degree of reliability as defaecography. Quantitative measures were not correlated, however, possibly because of methodological differences. The study confirms the value of the morphological data provided by 3DHRAM.
More Related Videos
03:49Anogenital Distance and Perineal Measurements of the Pelvic Organ Prolapse POP Quantification System
Published on: September 20, 2018
15:49Flexible Colonoscopy in Mice to Evaluate the Severity of Colitis and Colorectal Tumors Using a Validated Endoscopic Scoring System
Published on: October 16, 2013
Related Concept Videos
Assessment of the Rectum and Anus
Rectal Inspection
Begin by inspecting the perianal and anal areas for color, texture, rashes,...
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Imaging Studies VI: Voiding Cystourethrography and Cystography
Urodynamic Studies: Uroflowmetry