Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
EPID in vivo dosimetry in RapidArc technique
Krzysztof Slosarek1, Marta Szlag1, Barbara Bekman1
1Centre of Oncology - Maria Sklodowska Curie, Memorial Institute, Branch Gliwice, Radiotherapy and Brachytherapy Planning Department, Poland.
This study validates a method using an amorphous silicon electronic portal imaging device (aSi-EPID) for accurate in vivo dose estimation during RapidArc radiotherapy. The technique ensures precise dose delivery, crucial for effective cancer treatment.
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiotherapy Technology
Background:
- Accurate in vivo dose verification is critical in modern radiotherapy.
- The amorphous silicon electronic portal imaging device (aSi-EPID) offers potential for real-time dose monitoring.
- RapidArc is an advanced arc therapy technique requiring precise dose delivery.
Purpose of the Study:
- To estimate the delivered dose at the reference point using an aSi-EPID during patient treatment.
- To validate a novel method for in vivo dose evaluation in RapidArc radiotherapy.
Main Methods:
- A direct proportionality between EPID signal and delivered dose was assumed.
- Treatment plans were calculated for patients using the RapidArc technique.
- Dose was measured using an ionization chamber in a water-equivalent phantom and compared with EPID signal.
- An in vivo dose calculation formula was derived using EPID data and phantom measurements.
- Irradiation was performed on a Varian Clinac using an aSi-1000 EPID and Eclipse/Helios treatment planning system.
Main Results:
- In vivo dose calculations were performed for five patients.
- The planned dose at the reference point was 2 Gy for all treatment plans.
- The mean in vivo dose ranged from 1.96 Gy to 2.09 Gy, showing good agreement with planned doses.
Conclusions:
- The developed method is suitable for in vivo dose evaluation in RapidArc treatments.
- The aSi-EPID can be effectively utilized for accurate dose monitoring in clinical practice.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
Related Concept Videos
Absorption of Radiation
Radiation: Applications
The average...